| A
thick mix of Pulpdent Root Canal
Sealer is packed into the hub of the sterile needle. |
|
|
| The
needle hub is screwed into the Pressure
Syringe barrel. |
|
|
| Proper
placement of Pressure Syringe
needle 1-2 mm from apex. |
|
|
| One-quarter
turn of the screw plunger deposits sealer to the apex
first. Radiograph shows positive apical seal. The needle
is backed off to break contact with the sidewalls of the
canal, and the remainder of the canal is then back filled
to the orifice. |
|
|
| 1.
The needle should fit snugly 1-2 mm from the apex. |
 |
2.
Turn the screw plunger 1/4 turn clockwise. Wait
5 seconds. This fills the 1-2 mm at the apex. |
 |
| 3.
Back off needle to break contact with sidewalls.
Continue to turn the screw plunger until sealer
appears at the orifice. |
 |
4.
Using a cotton pliers, place a cotton pellet at
the orifice. Withdraw the needle while turning the
screw plunger. |
 |
|
|
Pulpdent
Root Canal Sealer - Case Studies
Resorbs with the Roots of
Deciduous Teeth
Permanent Filling Material
Tissue Compatible -
Slight Overfills often Resorb within 10-12 Months
|